Elemental Analysis XRF
Features
XGT-7200 Specification
Elements: Na to U
X-ray tube: Rh target / Tube voltage 50kV / Tube current 1mA
Fluorescent X-ray detector: Peltier cooled Silicon Drift Detector (SDD)
Transmitted X-ray detector: Nai(Tl) scintilltor
X-ray guide tube: Mono capillary 10μm / 100μm with no filter
Optical image: Full sample optical image and coaxial magnification image
Sample stage size: XY : 100mm×100mm
Sample chamber: Vacuum chamber model/ Max 300mm×300mm×80mm in vacuum
Signal processing: Digital pulse processor ( INCA unit )
Qualitative analysis: Auto identification / KLM marker / Peak search / Compare spectrum
Quantitative analysis: Non standard FPM / Standard FPM / Standard file matching FPM /Calibration curve / Multi layer FPM (Thickness gage) /Multi point analysis (Max5000) / Multi point result send to Excel / XGT-5000 spectrum view
Mapping function: Transmitted X-ray image / Elemental image / Spectrum mapping / Rectangle mapping / Generate spectrum / RGB composition / Scale maker / Line analysis